BACKGROUND: High-sensitivity cardiac troponin assays are being introduced clinically for earlier diagnosis of acute myocardial infarction (AMI). We evaluated the analytical performance of a high-sensitivity cardiac troponin T assay (hscTnT, Roche Diagnostics) in a multicenter, international trial. METHODS: Three US and 5 European sites evaluated hscTnT on the Modular® Analytics E170, cobas® 6000, Elecsys 2010, and cobas® e 411. Precision, accuracy, reportable range, an inter-laboratory comparison trial, and the 99th percentile of a reference population were assessed. RESULTS: Total imprecision (CVs) were 4.6-36.8% between 3.4 and 10.3 ng/L hscTnT. Assay linearity was up to 10,000 ng/L and the limit of blank and detection were 3 and 5 ng/L, respectively. The 99th percentile reference limit was 14.2 ng/L (n=533). No significant differences between specimen types, assay incubation time, or reagent lots existed. A substantial positive bias (76%) exists between the 4th generation and hscTnT assays at the low end of the measuring range (<50 ng/L). hscTnT serum pool concentrations were within 2SD limits of the mean of means in the comparison trial, indicating comparable results across multiple platforms and laboratories. CONCLUSION: The Roche hscTnT assay conforms to guideline precision requirements and will likely identify additional patients with myocardial injury suspicious for AMI.

Multicenter analytical evaluation of a high-sensitivity troponin T assay / A.K. Saenger, R. Beyrau, S. Braun, R. Cooray, A. Dolci, H. Freidank, E. Giannitsis, S. Gustafson, B. Handy, H. Katus, S.E. Melanson, M. Panteghini, P. Venge, M. Zorn, P. Jarolim, D. Bruton, J. Jarausch, A.S. Jaffe. - In: CLINICA CHIMICA ACTA. - ISSN 0009-8981. - 412:9-10(2011), pp. 748-754. [10.1016/j.cca.2010.12.034]

Multicenter analytical evaluation of a high-sensitivity troponin T assay

A. Dolci;M. Panteghini;
2011

Abstract

BACKGROUND: High-sensitivity cardiac troponin assays are being introduced clinically for earlier diagnosis of acute myocardial infarction (AMI). We evaluated the analytical performance of a high-sensitivity cardiac troponin T assay (hscTnT, Roche Diagnostics) in a multicenter, international trial. METHODS: Three US and 5 European sites evaluated hscTnT on the Modular® Analytics E170, cobas® 6000, Elecsys 2010, and cobas® e 411. Precision, accuracy, reportable range, an inter-laboratory comparison trial, and the 99th percentile of a reference population were assessed. RESULTS: Total imprecision (CVs) were 4.6-36.8% between 3.4 and 10.3 ng/L hscTnT. Assay linearity was up to 10,000 ng/L and the limit of blank and detection were 3 and 5 ng/L, respectively. The 99th percentile reference limit was 14.2 ng/L (n=533). No significant differences between specimen types, assay incubation time, or reagent lots existed. A substantial positive bias (76%) exists between the 4th generation and hscTnT assays at the low end of the measuring range (<50 ng/L). hscTnT serum pool concentrations were within 2SD limits of the mean of means in the comparison trial, indicating comparable results across multiple platforms and laboratories. CONCLUSION: The Roche hscTnT assay conforms to guideline precision requirements and will likely identify additional patients with myocardial injury suspicious for AMI.
Troponin ; 99th percentile ; High-sensitivity ; Acute myocardial infarction ; Cardiac biomarkers
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
2011
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/168976
Citazioni
  • ???jsp.display-item.citation.pmc??? 75
  • Scopus 318
  • ???jsp.display-item.citation.isi??? 300
social impact